Home > Products > Vacuum Measurement
Pirani Resistance Precision TransmitterPirani Resistance Precision TransmitterPirani Resistance Precision Transmitter

Pirani Resistance Precision Transmitter

    Pirani Resistance Precision Transmitter

The APG-500/APG500C precision resistance (Pirani) transmitter is designed for vacuum measurement in the range of 5×10−2 to 1.0×105 Pa.

This transmitter cannot be used to measure flammable or explosive gases that can react in certain atmospheres.

 

Technical Data

 

Measurement Principle: Based on the thermal conductivity physical properties of resistance (Pirani).

Gauge Interface: DN 16 ISO

Measurement Range: 5×10−2 to 1.0×105 Pa (only applicable to air, nitrogen, oxygen, carbon monoxide)

Accuracy (nitrogen):1.0×104 to 1.0×105 Pa: ±50% of reading

5.0×10−1 to 1.0×104 Pa: ±15% of reading

5.0×10−2 to 5.0×10−1Pa: ±50% of reading

Repeatability: 1×10−1 to 5×103 Pa: 5% of reading

Output Signal (Measurement Signal)

 

Voltage Range: 0 to +10.5 VDC

Output Voltage: +1.9 to +10.0 VDC

Voltage-Vacuum Ratio: Logarithmic 1.286 V/decade

Error Signal: 0 to +0.5 VDC (filament break)

Output Impedance: 2×5.1 Ω

Minimum Impedance: 10 kΩ

Response Time: 100 ms

Control Range: 3×10−1 to 5.0×103 Pa

Control Output: 1 relay contact output (DC30V/500mA)

Transmitter Identification: 27.0 kΩ

Adjustment: Two small holes on the transmitter side for full-scale (ATM atmospheric state) and zero-point (HV high vacuum state) adjustment

Power Supply

Transmitter Input Voltage: +13.5 to +30 VDC

Ripple: ≤5% Vpp

Current: <300 mA (**** startup current)

Power Consumption: ≤2 W

Cable Connection: FCC 68 / RJ45 electrical connector, 8-pin electrode

Transmitter Cable: 8-core cable

Cable Length: ≤100 meters (wire gauge not less than 8 × 0.18 square millimeters)

Grounding Concept: Refer to "Cable Connection"

Connection to Common Signal in Vacuum System: Through 10 kΩ resistor (voltage difference less than <45 V)

Separate Power and Signal Ground

Materials Exposed to Vacuum: Tungsten, nickel, stainless steel, nitrile rubber

 

Permissible Pressure: 0.3 MPa (absolute), limited to inert gases

 

Permissible Temperature:

 

Vacuum measurement: +5 to +45°C

Vacuum connection: 80°C

Sensor wire: 120°C

Storage: -5 to +60°C

Relative Humidity: Less than 80% when the temperature is ≤+31°C; reduced to 50% at +40°C

 

Usage: For indoor use only

 

Installation Direction: Vertical installation


Contact Us
  • E-mail: cysi@cysi.wang
  • Tel: +86 371 5519 9322
  • Fax: +86 371 8603 6875
  • Add: No. 820, 8th Floor, 1st Unit, 9th Block, Cuizhu Street, High-Tech Zone, Zhengzhou, Henan, China




Follow Us

Copyright © Zhengzhou CY Scientific Instrument Co., Ltd. All Rights Reserved    Update cookies preferences