Pirani Resistance Precision Transmitter
The APG-500/APG500C precision resistance (Pirani) transmitter is designed for vacuum measurement in the range of 5×10−2 to 1.0×105 Pa.
This transmitter cannot be used to measure flammable or explosive gases that can react in certain atmospheres.
Technical Data
Measurement Principle: Based on the thermal conductivity physical properties of resistance (Pirani).
Gauge Interface: DN 16 ISO
Measurement Range: 5×10−2 to 1.0×105 Pa (only applicable to air, nitrogen, oxygen, carbon monoxide)
Accuracy (nitrogen):1.0×104 to 1.0×105 Pa: ±50% of reading
5.0×10−1 to 1.0×104 Pa: ±15% of reading
5.0×10−2 to 5.0×10−1Pa: ±50% of reading
Repeatability: 1×10−1 to 5×103 Pa: 5% of reading
Output Signal (Measurement Signal)
Voltage Range: 0 to +10.5 VDC
Output Voltage: +1.9 to +10.0 VDC
Voltage-Vacuum Ratio: Logarithmic 1.286 V/decade
Error Signal: 0 to +0.5 VDC (filament break)
Output Impedance: 2×5.1 Ω
Minimum Impedance: 10 kΩ
Response Time: 100 ms
Control Range: 3×10−1 to 5.0×103 Pa
Control Output: 1 relay contact output (DC30V/500mA)
Transmitter Identification: 27.0 kΩ
Adjustment: Two small holes on the transmitter side for full-scale (ATM atmospheric state) and zero-point (HV high vacuum state) adjustment
Power Supply
Transmitter Input Voltage: +13.5 to +30 VDC
Ripple: ≤5% Vpp
Current: <300 mA (**** startup current)
Power Consumption: ≤2 W
Cable Connection: FCC 68 / RJ45 electrical connector, 8-pin electrode
Transmitter Cable: 8-core cable
Cable Length: ≤100 meters (wire gauge not less than 8 × 0.18 square millimeters)
Grounding Concept: Refer to "Cable Connection"
Connection to Common Signal in Vacuum System: Through 10 kΩ resistor (voltage difference less than <45 V)
Separate Power and Signal Ground
Materials Exposed to Vacuum: Tungsten, nickel, stainless steel, nitrile rubber
Permissible Pressure: 0.3 MPa (absolute), limited to inert gases
Permissible Temperature:
Vacuum measurement: +5 to +45°C
Vacuum connection: 80°C
Sensor wire: 120°C
Storage: -5 to +60°C
Relative Humidity: Less than 80% when the temperature is ≤+31°C; reduced to 50% at +40°C
Usage: For indoor use only
Installation Direction: Vertical installation