CY-ZJ-10B Medium Vacuum Hot Cathode Ionization Gauge
This type of gauge tube is a typical triode box structure, using a precious metal oxide cathode that is resistant to oxidation and can withstand instantaneous atmospheric impact. It is characterized by stable performance, long life, and is suitable for medium to high vacuum measurements.
Main Technical Parameters:
1.Measurement Range: 65 to 3×10−4Pa
2.X-ray Limit: 3×10−4Pa
3.Gauge Sensitivity: 0.015±15 Pa⁻¹
4.Operating Parameters:
Filament Voltage: 1.3 V
Filament Current: 2.5 A
Filament to Ground Potential: 50 V
Grid to Ground Potential: 165 V
Grid Current (Emission Current): 0.05/0.5 mA
Collector to Ground Potential: 0 V
5.Electron Bombardment (E.B) Degassing:
Grid Voltage: 300 V
Degassing Current: 20 mA
6.Vacuum Connection Types and Baking Temperature:
Φ15.5±0.5 Glass: 80°C
Φ15.5±0.1 Metal: 80°C
KF10/16 Quick-release Flange: 80°C
KF25 Quick-release Flange: 80°C
KF40 Quick-release Flange: 80°C
CF16 Ultra-high Flange: 450°C
CF25 Ultra-high Flange: 450°C
CF35 Ultra-high Flange: 450°C
Copyright © Zhengzhou CY Scientific Instrument Co., Ltd. All Rights Reserved Update cookies preferences